嫩草影院永久在线-人妻少妇被猛烈进入中文字幕-国精产品一区一区三区有限在线-国产免费一区二区三区在线观看-国产一区二区三区不卡在线观看

技術(shù)文章您現(xiàn)在的位置:首頁(yè) > 技術(shù)文章 > 人IL-6 ELISA KIT說明書

人IL-6 ELISA KIT說明書

更新時(shí)間:2011-08-18   點(diǎn)擊次數(shù):1664次

 

RD
Human Interleukin 6 (IL-6)

FOR RESEARCH USE ONLY
Assay range0.2 pg/ml -8 pg/ml               96determinations
Purpose
This kit allows for the determination ofIL-6concentrations in Humanserum, cellculture supernates and other biological fluids
 
Principle of the assay
The kit assay Human Interleukin 6(IL-6)level in the sampleuse Purified Human Interleukin 6(IL-6)antibody to coat microtiter plate wells, make solid-phase antibody, then addInterleukin 6(IL-6)to wells,Combined antibody which With HRP labeled goat anti- Human become antibody - antigen - enzyme-antibody complex, after washing Compley,Add TMB substrate solution,TMB substrate becomes blue color At HRP enzyme-catalyzed,reaction is terminated by the addition of a sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm. The concentration of  Human Interleukin 6(IL-6)in the samples is then determined by comparing the O.D. of the samples to the standard curve.
Materials provided with the kit

1
wash solution
20ml×1bottle
7
Stopp Solution
6ml×1 bottle
2
HRP-Conjugate reagent
6ml×1 bottle
8
Standard16pg/ml
0.5ml×1 bottle
3
Microelisa stripplate
12well×8strips
9
Standard diluent
1.5ml×1bottle
4
Sample diluent
6ml×1 bottle
10
Instruction
1
5
Chromogen Solution A
6ml×1 bottle
11
Closure plate membrane
2
6
Chromogen Solution B
6ml×1 bottle
12
Sealed bags
1

Specimen requirements
1.       extractas soon as possible after Specimen collection,and according to the relevant literature, and should be experiment as soon as possible after the extraction. If it can’t, specimen can be kept in -20 ℃ to preserve, Avoid repeated freeze-thaw cycles.
2.       Can’t detect the sample which contain NaN3, because NaN3 inhibits HRP active.
Assay procedure
1.       Dilute and add sample:Dilute Original density Standard as follow table:

8 pg/ml
5 Standard
150μl Original density Standard+150μl Standard diluent
4pg/ml
4 Standard
150μl 5 Standard+150μl Standard diluent
2 pg/ml
3 Standard
150μl 4 Standard+150μl Standard diluent
1 pg/ml
2 Standard
150μl 3 Standard +150μl Standard diluent
0.5 pg/ml
1 Standard
150μl 2 Standard +150μl Standard diluent

2.add sampleSet blank wells separay (blank comparison wells don’t add sample and HRP-Conjugate reagent, other each step operation is same). testing sample well. add Sample dilution 40μl to testing sample well, then add testing sample 10μl (sample final dilution is 5-fold), add sample to wells , don’t touch the well wall as far as possible, and Gently mix.
3.Incubate: After closing plate with Closure plate membrane ,incubate for 30 min at 37.
4.Configurate liquid: 30-fold wash solution diluted 30-fold (or 20-fold) with distilled water and reserve.
5.washingUncover Closure plate membrane, discardLiquid, dry by swing, add washing buffer to every well, still for 30s then drain,repeat 5 times, dry by pat.
6.add enzymeAdd HRP-Conjugate reagent 50μl to each well, except  blank well.
7.incubateOperation with 3.
8.washingOperation with 5.
9.colorAdd Chromogen Solution A 50ul and Chromogen Solution B to each well, evade the light preservation for 15 min at 37
10.Stop the reactionAdd Stop Solution50μl to each well, Stop the reaction(the blue color change to yellow color).
11.assaytake blank well as zero , Read absorbance at 450nm after Adding Stop Solution and within 15min.
Steps description

Standard, Sample diluent

 

AddStandard, Sample diluent, incubate for 30 min at 37.

 

Wash 5 time,AddHRP-Conjugate reagent, incubate for 30 min at 37.

 

Wash 5 times,Add Chromogen Solution A and B, incubate for 30 min at 37.

 

AddStopp Solution

 

Read absorbance at 450nm within 15 min

 

calculate

Calculate
Take the standard density as the horizontal, the OD value for the vertical ,draw the standard curve on graph paper, Find out the corresponding density according to the sample OD value by the Sample curve, multiplied by the dilution multiple, or calculate the straight line regression equation of the standard curve with the standard density and the OD value ,with the sample OD value in the equation, calculate the sample density, multiplied by the dilution factor, the result is the sample actual density.
Important notes
1.       The kit takes out from the refrigeration environment should be balanced 15-30 minutes in the room temperature, ELISA plates coated if has not use up after opened, the plate should be stored in Sealed bag.
2.       washing buffer will Crystallization separation, it can be heated the water helps dissolve when dilute . Washing does not affect the result.
3.       add Sample with sampler Each step, And proofread its accuracy frequently, avoids the experimental error. add sample within 5 min, if the number of sample is much , recommend to use Volley .
4.       if the testing material content is excessively higher (The sample OD is bigger than the first standard well ),please dilute Sample (n-fold), Please diluenteand multiplied by the dilution factor.×n×5.
5.       Closure plate membrane only limits the disposable use, to avoid cross-contamination.
6.       The substrate evade the light preservation.
7.       Please according to use instruction strictly, The test result determination must take the microtiter plate reader as a standard.
8.       All samples, washing buffer and each kind of reject should according to infective material process.
9.       Do not mix reagents with those from other lots.
 
Storage and validity
1Storage 2-8℃.
2validity six months.
 

聯(lián)